How to read an HPLC chromatogram on an analytical report

Most analytical reports include the chromatogram itself, not just the purity figure derived from it. This page walks through what each part of that plot shows.

Research-use scope. This page explains documents and analytical results. It is not for use in people or animals, contains no handling or procedural guidance, and does not replace the receiving laboratory's own acceptance criteria.

The plot

A chromatogram is a line graph. The horizontal axis is time in minutes from the moment the sample was introduced. The vertical axis is detector response, usually absorbance in milli-absorbance units at a stated wavelength. Each component of the sample reaches the detector at a characteristic time and shows up as a peak. The plot is the raw evidence behind the purity figure on the front page of the report.

Reading it left to right

  1. Solvent front. A small disturbance in the first minute or two is the unretained solvent passing the detector. Laboratories normally exclude it from integration. If it has been counted as a peak, the purity figure is being diluted by something that is not part of the sample.
  2. Baseline. The flat line between peaks. A stable, flat baseline means the instrument was equilibrated and the integration is trustworthy. A drifting or noisy baseline makes small peaks hard to separate from noise and is a reason to read the integration table with more caution.
  3. Main peak. The tallest peak, which should be the material named on the label. Its position on the time axis is the retention time. A symmetric, sharp peak indicates good chromatography; a shoulder or a split top suggests a closely eluting impurity that the method has not fully separated.
  4. Secondary peaks. Smaller peaks before or after the main one are other components: incomplete sequences, oxidation products, or process residues. Their retention times relative to the main peak are consistent from run to run under the same method, which is how a laboratory recognises a known impurity.
  5. Late region. Peaks well after the main one are more strongly retained components, sometimes aggregates or highly hydrophobic residues. A long tail of tiny peaks here is common and usually accounts for a fraction of a percent in total.

The integration table

Below or beside the plot, the report lists every integrated peak with its retention time, area, and area percentage. The purity figure is the area percentage of the main peak. Two checks make the table useful:

What to compare between lots

When two reports describe different lots of the same material under the same method, lay the chromatograms side by side. The main peak should sit at the same retention time within a few hundredths of a minute. The pattern of secondary peaks should look alike; a new peak that did not exist in the earlier lot is the most informative difference a chromatogram can show, more so than a small change in the headline percentage. See what a purity percentage means for how the figure itself is calculated.

FeatureLooks likeReading
Sharp, symmetric main peakNarrow triangle, single apexMethod resolved the main component cleanly
Shoulder on the main peakBump on one side of the apexClosely eluting impurity, incompletely separated
Flat baselineStraight line between peaksIntegration of small peaks is reliable
Rising baselineLine climbs across the runGradient or detector drift; small-peak areas are approximate
Unlabeled early spikeNarrow peak in the first minuteSolvent front; should be excluded from the calculation
On Arctic Lab Supply reports. The PDF linked from each entry in the batch COA library contains the laboratory's chromatogram and integration table for that lot, alongside the certificate number and the verification link. Each lot also has a batch page, for example BPC-157 5mg lot BCP-2502-B.

Related documentation