What a purity percentage means on a certificate of analysis

A purity figure is the number most people look at first on an analytical report. This page explains where it comes from, what it leaves out, and how to compare two of them fairly.

Research-use scope. This page explains documents and analytical results. It is not for use in people or animals, contains no handling or procedural guidance, and does not replace the receiving laboratory's own acceptance criteria.

Where the number comes from

On most research-material reports the purity figure is a chromatographic purity, produced by high-performance liquid chromatography (HPLC). The sample is dissolved, passed through a column, and its components leave the column at different times. A detector records a response over time, and each component appears as a peak.

The laboratory integrates the peaks and divides the area of the main peak by the total area of every peak it detected. A result of 99.2% means the main component accounted for 99.2% of the detected signal under that method. Nothing more, nothing less.

What the figure does not measure

Comparing two purity results

Two figures are only comparable when the method is comparable. Before treating 99.6% as better than 99.3%, check that:

  1. Both reports name the same method family, column type, and detection wavelength. A different wavelength can change which impurities are visible and therefore the denominator of the calculation.
  2. Both reports state a reporting limit or integration threshold. A method that ignores peaks under 0.1% will produce a higher purity than one that integrates everything down to 0.05%.
  3. Both describe the same material at the same labeled strength. A report for one strength does not describe another.
  4. Each report can be confirmed on the issuing laboratory's verification page. See how to verify a certificate of analysis.

With those points aligned, a difference of a few tenths of a percent is within what different runs of the same method routinely produce. A difference of several percent, or additional named impurity peaks, is worth reading closely.

Reading the impurity side of the report

The remainder after the main peak is the sum of everything else the detector saw. Better reports list the largest secondary peaks individually with their retention times and areas. A single unnamed impurity of 0.7% tells a different story from seven impurities of 0.1% each, even though both add up to the same purity figure. Truncated sequences and oxidation products typically elute close to the main peak; a well-resolved chromatogram shows them as separate, smaller peaks rather than a shoulder on the main one.

Statement on a reportWhat it establishesWhat it does not establish
Purity 99.4% (HPLC, area %)Main-peak share of detected signal under the stated methodIdentity, quantity, water or salt content
Observed mass 4113.5, theoretical 4113.6Identity of the main componentPurity or quantity
Net content 88%Share of vial mass that is the active materialChromatographic purity of that material
On Arctic Lab Supply reports. Each published report in the batch COA library shows the chromatographic purity as the laboratory reported it, with the method, test date, certificate number, PDF, and the issuing laboratory's verification link. Reports are lot-specific; for example BPC-157 5mg lot BCP-2502-B describes that lot only.

Related documentation